CH-FRU – Flux Processing
Contents
2023
FF-202401 (IRGA75)
- Final Flux Version: FF-202401
- Level-1 ID: Level-1_FR-20240228-000000_MERGED (combined results from Level-1_FR-20240203-162408 and Level-1_FR-20240227-140551_rECord)
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005 and Setup with rECord data logging, since 2023
- Instruments: R350, IRGA75
- Scripts: bico v1.6.0 (only for
sonicread
binary files for conversion to ASCII,rECord
files are already ASCII), fluxrun v1.4.0 (EddyPro v7.0.9), diive-legacy v0.21.0, diive v0.69.0 - FLUXNET upload: 28 Feb 2024
- Notes:
- Processing notes FF-202401: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
- Erroneous eddy covariance data between 12 Jun and 11 Jul 2023 14:00: EC data from the main tower during this time period must not be used. In June and July the sonic anemometer produced erroneous wind data, best seen in the clear offset period (in blue) for the unrotated wind component w, (see plot below). Luckily, we had parallel measurements from a test system (see below) in close proximity (metres) that can be used instead. The test system used a different, new logging script, see here.
- Parallel measurements from test system:
- This system was installed between 9 May 2023 and 11 Jul 2023 to test the new EC logging script
rECord
. Its EC raw data files are already logged in ASCII format, therefore a separate conversion to ASCII (viabico
) like we need for data recorded with thesonicread
script is not needed. Instead,rECord
data files can be used directly byfluxrun
andEddyPro
. - The time lag for the test system is longer than usual, with peak lag times of
10.2s
for the covariance between wind and CO2/H2O. This was due to a buffer implemented in therECord
script.
- This system was installed between 9 May 2023 and 11 Jul 2023 to test the new EC logging script
- The
rECord
files did not have AGC (window dirtiness) available as separate variable. AGC was therefore extracted from the gas analyzer diagnostic value and inserted as new column in all data files, usingdiive
v0.69.0. - Fluxes calculated from
sonicread
andrECord
files have overlapping time periods. During these periods,sonicread
fluxes was given preference.
CH-FRU 2023: In June and July the sonic anemometer produced erroneous wind data, best seen in the clear offset period (in blue) for the unrotated wind component w. EC data from this time period must not be used.
2022
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230425-205042
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
2021
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230425-205131
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
2020
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230425-205325
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-202101 (IRGA75, deprecated)
Final Flux Version: FF-202101Level-1 ID: Level-1_FR-20210418-162445Level-4 ID(s):Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: bico v0.5.0, fluxrun v0.5.0, diive-legacy v0.17.0FLUXNET upload: 19 Apr 2021Notes:19 Apr 2021 upload complete year, correction 25 May 2021
FF-202008 (IRGA75, deprecated)
Final Flux Version: FF-202008Level-1 ID: Level-1_ID2020-08-11T233528Level-4 ID(s):==check==
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT v0.9.5 with EP 7.0.6, Amp v0.16.0FLUXNET upload: 13 Aug 2020Notes:Time Period(s): 2020_1 (up until incl. 30 Jun 2020)Fluxes were re-calculated and now include all fluxes until 30 Jun 2020, needed for ICOS Winter 2020 Initiative.For more info see here: CH-MULTI / FF-202008.
FF-202005 (IRGA75, deprecated)
Final Flux Version: FF-202005Level-1 ID: Level-1_ID2020-06-01T144201Level-4 ID(s):none
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT v0.9.5 with EP 7.0.6FLUXNET upload: most likely no uploadNotes:Time Period(s): 2020_1 (up until incl. 12 May 2020)Data includes fluxes for 2020 so far (incl. 12 May 2020).For more info, see here: CH-FRU / FF-202005
2019
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_CH-FRU_FR-20230425-205432
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-202005 (IRGA75)
Final Flux Version: FF-202005Level-1 ID: Level-1_ID2020-05-31T135334Level-4 ID(s):Level-3-4_PI_FP2020_2005-2019Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT v0.9.5 with EP 7.0.6, Amp v0.16.0FLUXNET upload: 13 Aug 2020Notes:For more info, see here: CH-FRU / FF-202005
2018
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230425-205753
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-202005 (IRGA75, deprecated)
Final Flux Version: FF-202005Level-1 ID: Level-1_ID2020-05-29T103526Level-4 ID(s):Level-3-4_PI_FP2020_2005-2019Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT v0.9.5 with EP 7.0.6, Amp v0.16.0FLUXNET upload: 15 Aug 2020Notes:Updated fluxes: corrected for Wrong Calibration Gas 2017. For affected time periods see here: Setup since 2005.For more info, see here: CH-FRU / FF-202005
FF-201902 (IRGA75, deprecated)
Final Flux Version: FF-201902Level-1 ID: Level-1_ID2019-03-03T030626Level-4 ID(s):Level-3-4_FLUXNET2015-FN-20190606-beta-3
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT 0.86, EP 6.2.0 (Level-1)FLUXNET upload: 29 Mar 2019Notes:Known issues: Fluxes have not been corrected for the Wrong Calibration Gas 2017 issue (unknown at the time). See here for affected time periods: Setup since 2005.
2017
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230425-205849
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-202008 (IRGA75, deprecated)
Final Flux Version: FF-202008Level-1 ID: Level-1_ID2020-08-14T114135Level-4 ID(s):Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT v0.9.6 with EP 7.0.6, Amp v0.16.0FLUXNET upload: 15 Aug 2020Notes:Updated fluxes: corrected for Wrong Calibration Gas 2017. For affected time periods see here: Setup since 2005.For more info, see here: MULTI / FF-202008
FF-201902 (IRGA75, deprecated)
Final Flux Version: FF-201902Level-1 ID: Level-1_ID2019-03-03T030355Level-4 ID(s):Level-3-4_FLUXNET2015-FN-20190606-beta-3Level-3-4_PI_FP2020_2005-2019
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT 0.86, EP 6.2.0 (Level-1)FLUXNET upload: 29 Mar 2019Notes:Known issues: Fluxes have not been corrected for the Wrong Calibration Gas 2017 issue (unknown at the time). See here for affected time periods: Setup since 2005.There are erroneous flux data between 2017-06-09 00:30 and 2017-06-12 11:00 and then again between 2017-06-12 18:00 and 2017-06-16 11:00. Therefore, I manually set all qc_ flags for co2_flux, h2o_flux and LE to bad data (flag 2) in QC-20190310-150455 (Level-2). -LH
There are erroneous flux data between 2017-06-09 00:30 and 2017-06-12 11:00 and then again between 2017-06-12 18:00 and 2017-06-16 11:00. Therefore, I manually set all qc_ flags for co2_flux, h2o_flux and LE to bad data (flag 2) in QC-20190310-150455 (Level-2). -LH
2016
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230425-205953
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-201902 (IRGA75, deprecated)
Final Flux Version: FF-201902Level-1 ID: Level-1_ID2019-03-03T030144Level-4 ID(s):Level-3-4_FLUXNET2015-FN-20190606-beta-3Level-3-4_PI_FP2020_2005-2019Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT 0.86, EP 6.2.0 (Level-1)FLUXNET upload: 29 Mar 2019Notes:Known issues: Fluxes have not been corrected for the Wrong Calibration Gas 2017 issue (unknown at the time). See here for affected time periods: Setup since 2005. However, this issue only affects the last approx. 4 weeks of the year, a correction might therefore not be worthwhile since this has virtually zero impact on CO2 budgets etc.Level-1: The plotting of the full_output file was done in a separate step due to a min/max error during histogram creation in FCT 0.86. I simply added a try-except clause in the histogram code and that solved the problem (i.e. it is still FCT 0.86, but in this case with the clause, I did not give a new version number here because it only affects plotting). This clause will be added in future versions of FCT starting with FCT 0.88.
2015
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230425-210052
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-201606 (IRGA75, deprecated)
Final Flux Version: FF-201606Level-1 ID: Level-1_ID2016-07-02T110929Level-4 ID(s):Level-3-4_FLUXNET2015-FN-20190606-beta-3Level-3-4_PI_FP2020_2005-2019Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT 0.74, EP 6.1.0 (Level-1)FLUXNET upload: 4 Jul 2016Notes:June 2016: [2005-2015] v2016-06 calculated with EddyPro 610, FluxCalc 0.74, Anaconda3 v4
2014
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230425-210440
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-201606 (IRGA75, deprecated)
Final Flux Version: FF-201606Level-1 ID: Level-1_ID2016-07-02T110920Level-4 ID(s):Level-3-4_FLUXNET2015-FN-20161021Level-3-4_FLUXNET2015-FN-20190606-beta-3Level-3-4_PI_FP2020_2005-2019Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT 0.74, EP 6.1.0 (Level-1)FLUXNET upload: 4 Jul 2016Notes:June 2016: [2005-2015] v2016-06 calculated with EddyPro 610, FluxCalc 0.74, Anaconda3 v4
2013
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230425-210817
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-201606 (IRGA75, deprecated)
Final Flux Version: FF-201606Level-1 ID: Level-1_ID2016-07-02T110910Level-4 ID(s):Level-3-4_FLUXNET2015-FN-20161021Level-3-4_FLUXNET2015-FN-20190606-beta-3Level-3-4_PI_FP2020_2005-2019Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT 0.74, EP 6.1.0 (Level-1)FLUXNET upload: 4 Jul 2016Notes:June 2016: [2005-2015] v2016-06 calculated with EddyPro 610, FluxCalc 0.74, Anaconda3 v4Regarding QC: from 2013-06-24 until 2013-08-09: problems with AGC. AGC was not considered during this time period. AGC increased step-wise over a approx. 2-month period
CH-FRU FF-201606: From 2013-06-24 until 2013-08-09: problems with AGC. AGC was not considered during this time period. AGC increased step-wise over a approx. 2-month period
2012
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230425-220357
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-201606 (IRGA75, deprecated)
Final Flux Version: FF-201606Level-1 ID: Level-1_ID2016-07-02T110901Level-4 ID(s):Level-3-4_FLUXNET2015-FN-20161021Level-3-4_FLUXNET2015-FN-20190606-beta-3Level-3-4_PI_FP2020_2005-2019Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT 0.74, EP 6.1.0 (Level-1)FLUXNET upload: 4 Jul 2016Notes:June 2016: [2005-2015] v2016-06 calculated with EddyPro 610, FluxCalc 0.74, Anaconda3 v4
2011
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230425-220741
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-201606 (IRGA75, deprecated)
Final Flux Version: FF-201606Level-1 ID: Level-1_ID2016-06-30T213950Level-4 ID(s):Level-3-4_FLUXNET2015-FN-20161021Level-3-4_FLUXNET2015-FN-20190606-beta-3Level-3-4_PI_FP2020_2005-2019Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT 0.74, EP 6.1.0 (Level-1)FLUXNET upload: 4 Jul 2016Notes:June 2016: [2005-2015] v2016-06 calculated with EddyPro 610, FluxCalc 0.74, Anaconda3 v4
2010
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230425-221418
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-201606 (IRGA75, deprecated)
Final Flux Version: FF-201606Level-1 ID: Level-1_ID2016-06-30T213944Level-4 ID(s):Level-3-4_FLUXNET2015-FN-20161021Level-3-4_FLUXNET2015-FN-20190606-beta-3Level-3-4_PI_FP2020_2005-2019Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT 0.74, EP 6.1.0 (Level-1)FLUXNET upload: 4 Jul 2016Notes:June 2016: [2005-2015] v2016-06 calculated with EddyPro 610, FluxCalc 0.74, Anaconda3 v4Exception 2009-08-20 until 2010-03-09: problem with sonic w –> this problem results in problematic values for H. It is recommended to not use fluxes during this period. Flux values/flags/random error for CO2, H2O, LE, H and Tau were manually set to -9999 in the _QC_FINALFLAG file in folder 10_FINAL_QC of the respective year. Fluxes also set to -9999 in ghg-europe file. – LH 2016-07-02See plots below (2009)
2009
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230428-010024
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-201606 (IRGA75, deprecated)
Final Flux Version: FF-201606Level-1 ID: Level-1_ID2016-06-30T213938Level-4 ID(s):Level-3-4_FLUXNET2015-FN-20161021Level-3-4_FLUXNET2015-FN-20190606-beta-3Level-3-4_PI_FP2020_2005-2019Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT 0.74, EP 6.1.0 (Level-1)FLUXNET upload: 4 Jul 2016Notes:June 2016: [2005-2015] v2016-06 calculated with EddyPro 610, FluxCalc 0.74, Anaconda3 v4Exception 2009-08-20 until 2010-03-09: problem with sonic w –> this problem results in problematic values for H. It is recommended to not use fluxes during this period. Flux values/flags/random error for CO2, H2O, LE, H and Tau were manually set to -9999 in the _QC_FINALFLAG file in folder 10_FINAL_QC of the respective year. Fluxes also set to -9999 in ghg-europe file. – LH 2016-07-02
CH-FRU / FF-201606: Exception 2009-08-20 until 2010-03-09: problem with sonic w –> this problem results in problematic values for H. It is recommended to not use fluxes during this period. Flux values/flags/random error for CO2, H2O, LE, H and Tau were manually set to -9999 in the _QC_FINALFLAG file in folder 10_FINAL_QC of the respective year. Fluxes also set to -9999 in ghg-europe file.
CH-FRU / FF-201606: Exception 2009-08-20 until 2010-03-09: problem with sonic w –> this problem results in problematic values for H. It is recommended to not use fluxes during this period. Flux values/flags/random error for CO2, H2O, LE, H and Tau were manually set to -9999 in the _QC_FINALFLAG file in folder 10_FINAL_QC of the respective year. Fluxes also set to -9999 in ghg-europe file.
CH-FRU / FF-201606: Exception 2009-08-20 until 2010-03-09: problem with sonic w –> this problem results in problematic values for H. It is recommended to not use fluxes during this period. Flux values/flags/random error for CO2, H2O, LE, H and Tau were manually set to -9999 in the _QC_FINALFLAG file in folder 10_FINAL_QC of the respective year. Fluxes also set to -9999 in ghg-europe file.
2008
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230429-013552
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-201606 (IRGA75, deprecated)
Final Flux Version: FF-201606Level-1 ID: Level-1_ID2016-06-30T213932Level-4 ID(s):Level-3-4_FLUXNET2015-FN-20161021Level-3-4_FLUXNET2015-FN-20190606-beta-3Level-3-4_PI_FP2020_2005-2019Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT 0.74, EP 6.1.0 (Level-1)FLUXNET upload: 4 Jul 2016Notes: June 2016: [2005-2015] v2016-06 calculated with EddyPro 610, FluxCalc 0.74, Anaconda3 v4
2007
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230429-013855
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-201606 (IRGA75, deprecated)
Final Flux Version: FF-201606Level-1 ID: Level-1_ID2016-06-30T213928Level-4 ID(s):Level-3-4_FLUXNET2015-FN-20161021Level-3-4_FLUXNET2015-FN-20190606-beta-3Level-3-4_PI_FP2020_2005-2019Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT 0.74, EP 6.1.0 (Level-1)FLUXNET upload: 4 Jul 2016Notes: June 2016: [2005-2015] v2016-06 calculated with EddyPro 610, FluxCalc 0.74, Anaconda3 v4
2006
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230429-014233
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-201606 (IRGA75, deprecated)
Final Flux Version: FF-201606Level-1 ID: Level-1_ID2016-06-30T213921Level-4 ID(s):Level-3-4_FLUXNET2015-FN-20161021Level-3-4_FLUXNET2015-FN-20190606-beta-3Level-3-4_PI_FP2020_2005-2019Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT 0.74, EP 6.1.0 (Level-1)FLUXNET upload: 4 Jul 2016Notes: June 2016: [2005-2015] v2016-06 calculated with EddyPro 610, FluxCalc 0.74, Anaconda3 v4
2005
FF-202303 (IRGA75)
- Final Flux Version: FF-202303
- Level-1 ID: Level-1_FR-20230430-150230
- Level-4 ID(s):
- will be part of the next FLUXNET release in 2024
- Setup: Setup since 2005
- Instruments: R350, IRGA75
- Scripts: bico v1.3.0, fluxrun v1.2.0 (EddyPro v7.0.9), diive-legacy v0.21.0
- FLUXNET upload: 17 Sep 2023
- Notes:
- Processing notes FF-202303: Google sheet
- NO Angle-of-attack correction, cross-checked sonic orientation, cross-checked sonic height, no self-heating correction (not possible)
FF-201606 (IRGA75, deprecated)
Final Flux Version: FF-201606Level-1 ID: Level-1_ID2016-06-30T213909Level-4 ID(s):Level-3-4_FLUXNET2015-FN-20161021Level-3-4_FLUXNET2015-FN-20190606-beta-3Level-3-4_PI_FP2020_2005-2019Level-3-4_FLUXNET-WW2020_RELEASE-2022-1_FN-20220209
Setup: Setup since 2005Instruments: R350, IRGA75Scripts: FCT 0.74, EP 6.1.0 (Level-1)FLUXNET upload: 4 Jul 2016Notes: June 2016: [2005-2015] v2016-06 calculated with EddyPro 610, FluxCalc 0.74, Anaconda3 v4