CH-CHA: 2017

FF-202008
Setup | Setup since 2005 |
Time Period(s) | 2017_1, 2017_2, 2017_3 |
Instruments | R350, IRGA75 |
Final Flux Version | R350-IRGA75_FF-202008 |
Scripts | FCT v0.9.6 (Level-1 calcs), FQC v2.1.2 (Level-2 QC), Amp v0.16.0 |
Level-1 ID | Level-1_ID2020-08-06T141401 *** |
Level-2 ID | Level-2_QC-20200809-002012 |
Level-3-4 ID | soon |
Initiative | ICOS Winter 2020 Initiative (original upload 9 Aug 2020) |
Flux Calculations By |
LH |
Meteo Screening By |
IF |
Notes
- Fluxes 2017 were re-calculated because of the Wrong Calibration Gas 2017.
- For more info, see here: CH-CHA / FF-202008.
*** Level-1 adjusted *_full_output_*
file uploaded to EFDC for FLUXNET.
FF-201902
Setup | Setup since 2005 |
Time Period(s) | 2017_1 |
Instruments | R350, IRGA75 |
Final Flux Version |
R350-IRGA75_FF-201902 |
Scripts | FCT 0.86, EP 6.2.0 (Level-1), FM 0.91 (Level-1 merging), FQC 1.04 (Level-2) |
Level-1 ID | Level-1_ID2019-02-27T155352 / conversion in ID2019-02-12T172700 |
Level-2 ID | Level-2_QC-20190310-134532 * |
Level-3-4 ID | Level-3-4_FLUXNET2015-FN-20190606-beta-3 (Drought Study 2018) |
Initiative | European Drought Study 2018 (upload 29 Mar 2019) |
Flux Calculations By |
LH |
Meteo Screening By |
IF |
Notes
- Known issues: Fluxes have not been corrected for the Wrong Calibration Gas 2017 issue (unknown at the time). See here for affected time periods: Setup since 2005. Affected time period: all IRGA75 CO2 concentrations between
2017011114.C57
and2017031507.C00
(approx. 2 months).
* Level-2 uploaded to European Fluxes Database Cluster for FLUXNET data sharing